The objective of the current study was to develop a simple, accurate, precise and rapid RP-HPLC method and subsequently validate as per ICH guidelines for the determination of abacavir hydrochloride using mobile phase (mixture of Phosphate buffer pH 4.2 and acetonitrile in the ratio of 50:50) as the solvent. The proposed method involves the measurement of retention time at selected analytical wavelength. 218.0 nm was selected as the analytical wavelength. The retention time of abacavir HCl was found to be 2.824. The linearity of the proposed method was investigated in the range of 10-70 μg/ml (r2 = 0.998). The method was statistically validated for its linearity, accuracy and precision. Both inter-day and intra-day variation was found to be showing less than 2% RSD (relative standard deviation) value indicating high grade of precision of the method.